Yomi Robot's 92 FDA Adverse Event Reports Center on Implant Positioning Problems

Of the 92 reports the authors reviewed, 86 were filed with the FDA as injuries. Accuracy, calibration or positioning problems appeared in 57, and surgeons finished the case freehand in 30. The authors calculated no incidence or causal estimates.

Key TakeawayTwo researchers in Fujian, China, coded all 92 FDA MAUDE reports tied to Neocis's Yomi dental robot and found that 62 percent described accuracy, calibration or positioning problems. Implant revision or repositioning was the most common recovery step. The reports cannot show how often problems occur, since MAUDE carries no count of procedures.

Most adverse event reports filed with the FDA on Neocis's Yomi dental robot describe the implant ending up somewhere other than planned, according to an analysis of 92 reports in the Food and Drug Administration's MAUDE database published Nov. 18 in the Journal of Robotic Surgery.

Lin Wang of the Department of Product Design at Sanming University and Chenglu Ruan of the Department of Stomatology at Sanming Integrated Medicine Hospital, both in Fujian, China, searched MAUDE on Nov. 2, 2026, using the manufacturer name, the Yomi brand and FDA product codes PLV and QRY. Of the 92 reports they kept, 86 (93.5 percent) were filed as injuries, five as malfunctions and one without a type. Almost half the injury filings were thin on detail: 47 reports carried no affirmative patient-problem code.

Where the plan meets the patient

Each author read all 92 narratives separately against a fixed coding dictionary, and across 1,012 paired ratings in 11 categories they agreed every time. Accuracy, calibration or positioning signals appeared in 57 reports (62.0 percent). Problems with the robotic arm, the handpiece or interaction during surgery appeared in 30 (32.6 percent).

In 37 reports (40.2 percent) the implant was revised or repositioned, in 30 the surgeon completed the procedure freehand, and in 20 (21.7 percent) the team placed a graft or membrane. The authors wrote that the reports identify recurring failure modes "particularly around plan-to-patient alignment and patient-interface management."

Because one manufacturer report can describe the same event twice, the pair tested for duplicates. Collapsing one text-identical pair left 91 units, and merging six conservative candidate clusters left 84. The positioning share held at 61.5 percent and 61.9 percent. Only 46 reports listed the date the manufacturer received the complaint.

The reports, the authors wrote, "do not establish incidence, comparative safety, confirmed device malfunction, or causality." They declared no competing interests and disclosed that a large language model helped with literature discovery, language editing and early organization of the coding rules, with both authors verifying the classifications and numbers. The full paper sits behind Springer's paywall, and this account draws on the published abstract and declarations.

A decade of 510(k) clearances

The FDA's 510(k) database lists 19 clearances for Neocis, starting with the Neocis Guidance System on Dec. 22, 2016. The most recent, K261321 for the Yomi S Robotic System, was decided Oct. 13, 2026, under the dental navigation system product code QRY with PLV, the dental stereotaxic instrument code, listed as a subsequent code. The Miami company's first clearance under QRY came in November 2022, and earlier ones were listed under PLV.

The 92 Reports in Numbers
  • Filed as injury 86, malfunction 5, unspecified 1
  • Accuracy, calibration or positioning signal: 57
  • Arm, handpiece or intraoperative interaction signal: 30
  • Revision or repositioning 37, freehand completion 30, graft or membrane 20

Why This Matters to Oral Surgeons

Reports naming Neocis peaked in 2022, when openFDA logs 39 received, then fell to 22 in 2023, 10 in 2024 and six in 2025. MAUDE cannot say whether that slide tracks fewer problems, fewer robotic cases or a change in who files, and the Yomi S cleared in October has no reporting history at all. The authors propose verification, stopping, conversion and interface-release steps for robotic placement and say they should be tested prospectively, which a new hardware generation entering practice now would allow.

Sources

  1. Wang L, Ruan C, "Post-market safety signals of the Yomi dental robotic system: an analysis of FDA MAUDE reports," Journal of Robotic Surgery 20(1):1228, Nov. 18, 2026 (PMID 43050053). pubmed.ncbi.nlm.nih.gov
  2. U.S. Food and Drug Administration, 510(k) Premarket Notification K261321, Yomi S Robotic System, decided Oct. 13, 2026. accessdata.fda.gov